1 Scope
This part of IEC 60565 specifies methods and procedures for free-field calibration of hydrophones, as well as individual electroacoustic transducers that can be used as hydrophones (receivers) and/or projectors (source transducers). Two general types of calibration are covered within this document: absolute calibration
using the method of three-transducer spherical-wave reciprocity, and relative calibration by comparison with a reference
device which has already been the subject of an absolute calibration.
The maximum frequency range of the methods specified in this document is from 200 Hz
to 1 MHz. The lowest acoustic frequency of application will depend on a number of
factors, and will typically be in the range 200 Hz to 5 kHz depending mainly on the
dimensions of the chosen test facility, The highest frequency of application for the
methods described here is 1 MHz.
Procedures for pressure hydrophone calibration at low frequencies can be found in IEC 60565‑2 [1] 1. Procedures for hydrophone calibration at acoustic frequencies greater than 1 MHz are covered by IEC 62127‑2 [2].
Excluded from the scope of this document are low-frequency pressure calibrations of
hydrophones, which are described in IEC 60565‑2 [1]. Also excluded are calibrations of digital hydrophones and systems, calibration of marine autonomous acoustic recorders, calibration of
acoustic vector sensors such as particle velocity sensors and pressure gradient hydrophones, calibration of passive sonar arrays consisting of multiple hydrophones, and calibration of active sonar arrays consisting of projectors and hydrophones.
This document presents a description of the requirements for free-field calibration
in terms of test facility, equipment and instrumentation, signal processing, and frequency limitations. A description of achievable uncertainty and
rules for the presentation of the calibration data are provided. Also included are
informative annexes that provide additional guidance on
-
• measurement of directional response of a hydrophone or projector,
-
• measurement of electrical impedance of hydrophones and projectors,
-
• electrical loading corrections,
-
• acoustic far-field criteria in underwater acoustic calibration,
-
• pulsed techniques in free-field calibrations,
-
• assessment of uncertainty in the free-field calibration of hydrophones and projectors,
-
• derivation of the formulae for three-transducer spherical-wave reciprocity calibrations,
-
• calibration using travelling-wave tubes,
-
• calibration of hydrophones using optical interferometry, and
-
• calibrations in reverberant water tanks using continuous signals.