1 Scope
This document specifies a method for calibrating the kinetic energy scales of Auger
electron spectrometers with an uncertainty of 3 eV, for general analytical use in
identifying elements at surfaces. This document also specifies a method for establishing
a calibration schedule.
It is applicable to instruments used in either direct or differential mode, where
the resolution is less than or equal to 0,5 % and the modulation amplitude for the
differential mode, if used, is 2 eV peak-to-peak. It is applicable to those spectrometers
equipped with an inert gas ion gun or other method for sample cleaning and with an
electron gun capable of operating at 4 keV or higher beam energy.