1. Scope
1.1 This practice covers the procedure for performing corrosion fatigue tests to obtain
S-N (3.2.1) fatigue curves or statistically derived fatigue strength values, or both, for metallic
implant materials. This practice describes the testing of axially loaded fatigue specimens
subjected to a constant amplitude, periodic forcing function in saline solution at
37°C and in air at room temperature. The environmental test method for implant materials
may be adapted to other modes of fatigue loading such as bending or torsion. While
this practice is not intended to apply to fatigue tests on implantable components
or devices, it does provide guidelines for fatigue tests with standard specimens in
an environment related to physiological conditions.