1. Scope
1.1 This practice provides experimental protocols for bio-logical assays of tissue reactions
to absorbable biomaterials for implant applications. This practice applies only to
absorbable materials with projected clinical applications in which the materials will
reside in bone or soft tissue longer than 30 days and less than three years. Other
standards with designated implantation times are available to address shorter time
periods. Careful consideration should be given to the appropriateness of this practice
for slowly degrading materials that will remain for longer than three years. It is
anticipated that the tissue response to degrading biomaterials will be different from
the response to nonabsorbable materials. In many cases, a chronic inflammatory response
may be observed during the degradation phase, but the local histology should return
to normal after absorption; therefore, the minimal tissue response usually equated
with biocompatibility may require long implantations.