1. Scope
1.1 This guide describes a laboratory method using a weight-loss technique for evaluating
the wear properties of materials or devices, or both, which are being considered for
use as bearing surfaces of human-hip-joint replacement prostheses. The hip prostheses
are evaluated in a device intended to simulate the tribological conditions encountered
in the human hip joint, for example, use of a fluid such as bovine serum, or equivalent
pseudosynovial fluid shown to simulate similar wear mechanisms and debris generation
as found in vivo, and test frequencies of 1 Hz or less.