1 Scope
This International Standard specifies methods for the determination of pore water pressure in both unsaturated and saturated soil using tensiometers. The methods are applicable
for in situpore water pressure measurements in the field, as well as for monitoring pore water pressure in, for example, plant containers or soil cores used in experimental procedures.
At normal atmospheric pressures, i.e. about 100 kPa, the application of these methods
is limited to a range of pressures down to about – 85 kPa. The range is reduced at
lower atmospheric pressures. Tensiometers will not function if sub-zero temperatures
occur at the measurement depth. Their accuracy is influenced by soil and air temperature
fluctuations. Tensiometer response time ranges from a few seconds to several days.
To obtain reliable measurements under field conditions, tensiometers require frequent
servicing.
A tensiometer provides point measurements of pore water pressure. To measure pore water pressure at different depths, several tensiometers will be necessary. In the field, replicate
sets of instruments will be required if the spatial variability of the soil is to
be allowed for.