1 Scope and object
This clause of Part 1 is applicable, except as follows:
1.1.1 Equipment included in scope
Replacement:
Replace the second paragraph by the following:
This Part 2 of IEC 61010 specifies particular safety
requirements for the following types a) to c) of electrical equipment and their
accessories, wherever they are intended to be used, whenever that equipment
incorporates refrigerating systems as an integral part of, or
separate from, the equipment and the equipment is in direct control of the
refrigerating system.
This document details all the requirements when up to 150 g of flammable
refrigerant are used per stage of a
refrigerating system. Additional requirements
beyond the current scope of this document apply if a refrigerant charge of flammable
refrigerant exceeds this amount.
Addition:
Add the following text after the last paragraph:
NOTE 101 Examples for refrigerating equipment include, but are not limited to, laboratory equipment such as laboratory refrigerators, freezers, refrigerated display cabinets.
It is possible that all or part of the equipment falls within the scope of one or
more other Part 2 standards of IEC 61010 as well as within the scope of this
standard. In that case, the requirements of those other Part 2 standards will
also apply. In particular, if equipment is intended to be used as a centrifuge,
the requirements of IEC 61010‑2‑020 apply. However, when the
equipment incorporates a refrigerating system and a heating
function where the combination of the two introduces additional or more severe
hazards than if treated separately, then it is possible that IEC 61010‑2‑012 is applicable instead of this
Part 2-011.
See further information in the flow chart (Figure 102) for the selection process and guidance in the
Introduction.