1 Scope and object
1.1.1 Equipment included in scope
NOTE 1 This part includes also the requirements of Part 2-030. Testing and measuring circuits that are not within the scope of this part are considered to be covered by the requirements of Part 1 or other parts 2s of IEC 61010, and then will also need to meet the requirements of these other parts with the exception of Part 2-030. Current clamp meters and similar currents sensors that have a primary purpose of measuring voltage on a live mains circuit are also within the scope of Part 2-033.
NOTE 2 Some current sensors are also known as current clamps and current probes.
NOTE 3 Some current sensors designed for portable use can also be used for fixed installations.
-
a) Type A: a current sensor designed to be applied around or removed from uninsulated hazardous live conductors. Type A current sensors have defined hand-held or hand-manipulated parts providing protection against electric shock from the conductor being measured, and also have protection against short-circuits between wires and busbars during clamping.
-
b) Type B: a current sensor which has protection against short-circuits between wires or busbars during clamping but without defined hand-held or hand-manipulated parts which provide protection against electric shock during clamping. Additional protective means are necessary to avoid electric shock from hazardous live conductors which cannot be de-energised during application or removal of the current sensor.
EXAMPLE 1
Flexible current sensors. -
c) Type C: a current sensor without protection against short-circuits between wires or busbars during clamping. Type C current sensors are intended to be applied to or removed from uninsulated hazardous live conductors or from non-limited-energy circuit conductors only when they are de-energised.
EXAMPLE 2
Split-core transducers. -
d) Type D: a current sensor designed to be applied around or removed from insulated conductors or from limited-energy circuit conductors.A Type D current sensor does not need protection against short-circuits during clamping and has no defined hand-held or hand-manipulated parts providing protection against electric shock from the conductor being measured.
EXAMPLE 3
Current probes for oscilloscopes and earth leakage current detectors.
NOTE 4 All current sensors can also be used around insulated conductors. In this case, hazards are limited to acceptable levels by the insulation of the conductors.