1 Scope
This European Standard specifies a procedure on how to improve the performance of
a routine XRF method, already in use for analysis of high alloy steels, by using a
”near by technique”.
The ”near by technique” requires at least one target sample (preferable a CRM) of
a similar composition as the unknown sample.
The method is applicable to elements within the concentration ranges according to
Table 1:
Table 1
—
Concentration ranges
| Element | Concentration range, % (m/m)a |
|---|---|
| Si | 0,05 to 1,5 |
| Mn | 0,05 to 5,0 |
| P | 0,005 to 0,035 |
| Cr | 10 to 25 |
| Ni | 0,1 to 30 |
| Mo | 0,1 to 6,5 |
| Cu | 0,02 to 1,5 |
| Co | 0,015 to 0,30 |
| V | 0,015 to 0,15 |
| Ti | 0,015 to 0,50 |
| Nb | 0,05 to 1,0 |
The method is applicable to analysis of either chill-cast or wrought samples having
a diameter of at least 25 mm and with a carbon concentration of less than 0,3 % (see
NOTE). Other elements should have a concentration below 0,2 %.
NOTE High carbon concentrations, in combination with high Mo and Cr concentrations, could have undesirable structural effects on the sample and could affect the determination of phosphorus and chromium, in particular.
Matrix effects exist between the elements listed. To compensate for those inter-element
effects, mathematical corrections shall be applied. A variety of computer programs
for corrections is commonly used and included in the software package from the manufacturers.