1 Scope
This standard applies to all kinds of passive dosimetry systems that are used for measuring
-
— the personal dose equivalentHp(10) (for whole body dosimetry),
-
— the personal dose equivalentHp(3) (for eye lens dosimetry),
-
— the personal dose equivalentHp(0,07) (for both whole body and extremity dosimetry),
-
— the ambient dose equivalent H*(10) (for environmental dosimetry), or
-
— the directional dose equivalentH’(0,07) (for environmental dosimetry).
Text deleted
This standard applies to dosimetry systems that measure external photon and/or beta radiation in the dose range between 0,01 mSv
and 10 Sv and in the energy ranges given in Table 1. All the energy values are mean energies with respect to the prevailing dose quantity.
The dosimetry systems usually use electronic devices for the data evaluation and thus are often computer
controlled.
Table 1
—
Mandatory and maximum energy ranges covered by this standard
| Measuring quantity | Mandatory energy range for photon radiation | Maximum energy range for testing photon radiation | Mandatory energy range for beta-particle radiationa | Maximum energy range for testing beta-particle radiationa |
|---|---|---|---|---|
|
H
p(10),
H*(10)
|
80 keV to 1,25 MeV | 12 keV to 10 MeV | — | — |
| H p(3) | 30 keV to 250 keV | 8 keV to 10 MeV | 0,8 MeV almost equivalent to an Emax of 2,27 MeV | 0,7 MeVb to 1,2 MeV almost equivalent to Emax from 2,27 MeV to 3,54 MeV |
|
H
p(0,07),
H’(0,07)
|
30 keV to 250 keV | 8 keV to 10 MeV | 0,8 MeV almost equivalent to an Emax of 2,27 MeV | 0,06 MeVc to 1,2 MeV almost equivalent to Emax from 0,225 MeV to 3,54 MeV |
a The following beta radiation source are suggested for the different mean energies:
For 0,06 MeV: 147Pm; for 0,8 MeV: 90Sr/90Y; for 1,2 Mev: 106Ru/106Rh.
NOTE 2 In this standard, “dose” means dose equivalent, unless otherwise stated.
NOTE 3 For Hp(10) and H*(10) no beta radiation is considered. Reasons: 1) Hp(10) and H*(10) are a conservative estimate for the effective dose which is not a suitable quantity for beta radiation. 2) No conversion coefficients are available in ICRU 56, ICRU 57 or ISO 6980‑3.
NOTE 4 The maximum energy ranges are the energy limits within which type tests according to this standard are possible.
The test methods concerning the design (Clause 8), the instruction manual (Clause 9), the software (Clause 10), environmental influences (Clause 13), electromagnetic influences (Clause 14), mechanical influences (Clause 15), and the documentation (Clause 16) are independent of the type of radiation. Therefore, they can also be applied to
other dosimetry systems, e.g. for neutrons, utilizing the corresponding type of radiation for testing.
This standard is intended to be applied to dosimetry systems that are capable of evaluating doses in the required quantity and unit (Sv) from
readoutsignals in any quantity and unit. The only correction that may be applied to the evaluated
dose (indicated value) is the one resulting from natural background radiation using extra dosemeters.
NOTE 5 The correction due to natural background can be made before or after the dose calculation.