1 Scope
This document applies to all types of dosemeters, irrespective of the type of radiation
intended to be measured. Tests according to this document determine whether a single
radiation pulse can be measured correctly even if the dosemeter is in the internal state relevant for measuring background or environmental radiation. The characteristics
of the dosemeter for repeated pulses is expected to be better than for one single
radiation pulse with the same parameters but worse than for continuous radiation, i.e., in between of the characteristics for these two extreme conditions.
The pulsed radiation source is characterized by the parameters:
-
– the maximum dose rate during the radiation pulse, , occurring at the workplace,
-
– the maximum dose per radiation pulse, Hpulse,max, occurring at the workplace,
-
– the minimum radiation pulse duration, tpulse,min, occurring at the workplace, and
-
– the range of the pulse repetition frequency, fpulse, occurring at the workplace.
Annex A gives some parameter values for typical workplaces where pulsed radiation occurs.
This document considers the pulsation of the radiation field as an additional influence
quantity like particle energy and direction of radiation incidence. Therefore, the
tests described are additional to all the tests in the instrument specific standards.
This document describes methods to determine the following characteristic parameters
of the dosemeters:
-
– the maximum measurable dose rate in the pulse, ,
-
– the maximum measurable dose in the pulse, Hmeas,max,
-
– the minimal pulse duration, tmeas,min, and
-
– the range for the pulse repetition frequency, fmeas,min to fmeas,max.
NOTE These parameters may be inter-related depending on the detector used.
It is applicable to photon radiation but basically can be adapted to all types of
radiation for which a suitable pulsed reference field is available. The term dose
is used in this document in the sense of dose equivalent, but the requirements can
also be adapted to air kerma, exposure or other quantities expressing the amount of
radiation.
The parameter pulse repetition frequency, fpulse, is included in the testing procedures, but for this inclusion additional work has
to be done. Especially, reference fields for radiation conditions in surrounding fields
of accelerators are missing (high pulse repetition frequency, ultra-short pulses).