1 Scope
This document establishes the currently recognized approaches and special considerations
needed when evaluating the in vitro and in vivo performance of absorbable metals and
implants fabricated, in whole or in part, from them. This document describes how the evaluation
of these metals can differ from those utilized for permanent non-absorbable implantable
implants (or subcomponents), in that absorbable metal implants (or subcomponents) are – by design – intended to be absorbed in their entirety by
the host.
This document provides guidance regarding the materials considerations, in vitro degradation/fatigue
characterization, and biological evaluation of medical implants made of absorbable metals. The provided content is intended to deliver added clarity
to the evaluation of these materials and implants to increase awareness of critical factors and reduce potential for generation of
erroneous or misleading test results.
While this document and the herein described referenced standards contain many suggested
alterations or modifications to currently practiced procedures or specifications,
the provided content is intended to complement, and not replace, current conventions
regarding the assessment of implantable implants.
This document covers the evaluation of absorbable metal specific attributes in general
and is not intended to cover application or implant specific considerations. Thus, it is important to consult relevant implant and/or application specific standards.
This document does not apply to non-absorbable or non-metallic components (e.g. polymeric
coatings, pharmaceuticals, non-absorbable metals) used in conjunction with absorbable
metal implants.