1 Scope
This part of IEC 62209 series is applicable to any wireless communication device capable of transmitting electromagnetic
fields (EMF) intended to be used at a position near the human body, in the manner
described by the manufacturer, with the radiating part(s) of the device at distances
up to and including 200 mm from a human body, i.e. when held in the hand or in front
of the face, mounted on the body, combined with other transmitting or non-transmitting
devices or accessories (e.g. belt-clip, camera or Bluetooth add-on), or embedded in garments. For transmitters
used in close proximity to the human ear, the procedures of IEC 62209‑1:2005 are applicable.
This standard is applicable for radio frequency exposure in the frequency range of
30 MHz to 6 GHz, and may be used to measure simultaneous exposures from multiple radio
sources used in close proximity to human body. Definitions and evaluation procedures
are provided for the following general categories of device types: body-mounted, body-supported,
desktop, front-of-face, hand-held, laptop, limb-mounted, multi-band, push-to-talk,
clothing-integrated. The types of devices considered include but are not limited to
mobile telephones, cordless microphones, auxiliary broadcast devices and radio transmitters
in personal computers.
This International Standard gives guidelines for a reproducible and conservative measurement
methodology for determining the compliance of wireless devices with the SAR limits.
Because studies suggest that exclusion of features to represent a hand in human models
constitutes a conservative case scenario for SAR in the trunk and the head, a representation
of a hand is not included if the device is intended to be used next to the head or
supported on or near the torso [73], [80]. This standard does not apply for exposures from transmitting or non-transmitting
implanted medical devices. This standard does not apply for exposure from devices
at distances greater than 200 mm away from the human body.
IEC 62209‑2
makes cross-reference to IEC 62209‑1:2005 where complete clauses or subclauses apply, along with any changes specified.