1 Scope
This Technical Specification specifies an inductively coupled plasma optical emission
spectroscopic (ICP-OES) method for the determination of trace element content in oils. Depending on the dilution solvent used, most types of vegetable
oils can be analysed (crude, degummed, refined, bleached, deodorized and hardened
oils) and nearly all types of lecithins and phosphatides.
This procedure is only suitable when the elements are present in a solubilized form.
Fine particles, such as bleaching earth, metal particles and rust, can result in poor
recovery of the trace elements present as nebulization and atomization problems affect the ICP-OES analysis.
NOTE The only suitable non-ashing direct method for samples containing fine particles is graphite furnace atomic absorption spectrometry.