1 Scope
This International Standard provides essential performance and safety requirements
for oro-tracheal and naso-tracheal tubes and tracheal tube connectors. Tracheal tubes with walls reinforced with metal or nylon, tracheal tubes with shoulders, tapered tracheal tubes, tracheal tubes with means for suctioning, monitoring or delivery of drugs or other gases, and the
many other types of tracheal tubes devised for specialized applications are included in this International Standard,
as many specialized tracheal tubes are now commonly used, and all share similar essential requirements as defined in
this International Standard.
Endobronchial (including tracheobronchial) tubes, tracheostomy tubes, and supralaryngeal
airways are excluded from the scope of this International Standard.
Tracheal tubes
intended for use with flammable anaesthetic gases or agents, lasers, or electrosurgical
equipment are outside the scope of this International Standard.
NOTE References [1] to [4] deal with laser surgery of the airway.