1 Scope
This part of ISO 11904 specifies basic framework measurement methods for sound immission from sound sources
placed close to the ear. These measurements are carried out with a manikin, equipped
with ear simulators including microphones. The measured values are subsequently converted
into corresponding free-field or diffuse-field levels. The results are given as free-field
related or diffuse-field related equivalent continuous A-weighted sound pressure levels.
The technique is denoted the manikin technique.
This part of ISO 11904 is applicable to exposure to sound from sources close to the ear, for example during
equipment tests or at the workplace to sound from earphones or hearing protectors
with audiocommunication facilities.
This part of ISO 11904 is applicable in the frequency range from 20 Hz to 10 kHz. For frequencies above
10 kHz, ISO 11904‑1 can be used.